Pacingappnotenew.fm
Non-Invasive Transcutaneous Pacing
Application Note
This Application Note provides an overview of non-invasive transcutaneous pacing. It includes a brief history of pacing, describes the electrical activity of the heart, and explains how pacing works.
Non-Invasive Transcutaneous Pacing
What is Non-Invasive Transcutaneous Pacing?
Non-invasive transcutaneous pacing is a technique of
cardiac depolarization and myocardial contraction.
electrically stimulating the heart externally through a set
Pacing is one method of treating patients when their
of electrode pads. The stimulus is intended to cause
heart's own conduction system slows dangerously.
Electrical Activity of the Heart
The unique physiologic characteristics of cardiac muscle
In order for the heart to beat with a smooth and efficient
cells produce intrinsic pacing and efficient conduction,
pumping action, the various parts of the myocardium
leading to optimal synchrony and contractility (see
must contract in a well-defined sequence. The atria must
contract before the ventricles, the ventricles must begin contracting near the apex, etc. An orderly sequence of
The Cardiac Cycle 1
contraction during systole is provided for by specialized conductive pathways in the myocardium.
These pathways consist of highly specialized myocardial
cells which have the ability to conduct a depolarization
wave at a much greater velocity than ordinary myocardial
The normal sequence of myocardial depolarization and
contraction, is described below:
Ventricular systole
A cluster of highly conductive cells called the
sino-atrial (SA) node is located near the back wall of the right atrium at the SA node that the depolarization wave is normally initiated.
Approximate time (sec)
Normal Myocardial Sequence
The myocardial cell membrane, like the membranes of other muscle cells in the body, has the ability to conduct
a propagated action potential or depolarization wave. Likewise, depolarization of the myocardial cell membrane by a propagated action potential results in myocardial contraction. The propagation of an action potential can be initiated by intrinsic automaticity, direct electrical stimulation (i.e. transvenous or transthoracic pacing), or by a transmitted external stimulus.
The cardiac cell membrane does not require an external stimulus to reach threshold. It possesses a special ability
that allows it to spontaneously depolarize at periodic
From the SA node, the depolarization wave passes
intervals without the need for an external stimulus. This
from right to left over both atria, resulting in atrial
property is called automaticity.
systole. Within 70 ms, all portions of both atria have started to contract.
Electrical Activity of the Heart
The atrio-ventricular (AV) node consists of a cluster
Because our bodies require more oxygen during periods
of cells leading from the lower portion of the right
of physical and emotional stress, the SA node is liberally
atrium to the ventricular septumhe
supplied with nerve endings which can stimulate its cells
AV node is highly specialized and conducts the
more rapidly when necessary. Hence, during periods of
depolarization wave very slowly. It delays its progress
excitement, the SA node responds with a faster rate of
for approximately 70 ms, which allows atrial systole
depolarization and the heart rate increases to ensure the
to reach completion before ventricular systole begins.
muscles have enough oxygen to cope with the sudden demand.
The properly functioning Bundle of His and Purkinje network ensure that the heart always contracts rhythmically, and therefore pumps efficiently. The fact that all conductive cells in the heart possess the capability of spontaneous depolarization provides a natural back-up pacemaking capability, in case the SA node should fail.
The progress of the depolarization wave causing the muscle contraction can be followed and recorded. The recorded ECG waveforms and their correlation to the activity of the heart are illustrated in Figureh
The depolarization of the atria is represented by the P-wave). During this time, the atria contract and force blood into the ventricles. At the point labeled "Q", the ventricles begin depolarizing and contracting.
From the AV node, the depolarization wave moves to
the Bundle of His and its bundle branches. These lie
Atrial Depolarization
in the ventricular septum and conduct the
depolarization wave to the apex of the heart.
From the bundle branches, the depolarization wave
rapidly travels through the Purkinje network, a fine
mesh of highly conductive fibers that cover the endocardial (inner) surfaces of both ventricles.
At this point, depolarization and contraction of the
ventricular myocardium begin. From the Purkinje network, the depolarization wave moves through the ventricular myocardium from the endocardial (inner) to the epicardial (outer) surfaces. Because of its course of travel through the bundle branches and Purkinje network, myocardial contraction begins in the ventricular septum, then moves rapidly to the apex of the heart and finally proceeds towards the base.
As soon as the cells in the myocardium have
depolarized, they begin entering the refractory state and repolarizing. They rest until the next stimulus, which under normal conditions, is supplied by the cells in the SA node.
Non-Invasive Transcutaneous Pacing
Notice the time delay between the P-wave and Q. This is
The "T" wave represents the refractory period of the
the 70 ms delay introduced by the AV Node).
heart when the ventricles are repolarizing ).
AV Node Delay
Ventricular Repolarization
The time between "Q" and "S" (called the QRS complex) immediately precedes full contraction of the ventricles when blood is forced into the arterial system .
Ventricular Depolarization
Application of Pacing
Disease in the conduction system or failure in the
symptomatic bradycardia or other arrhythmias which
automaticity of the myocardial cells can lead to
may necessitate pacing intervention.
Application of Pacing
The following table explains the indications and
Fixed Rate or Demand
symptoms of emergency pacing.
Pacing—Which One To Use?
Indications of Emergency Pacing and
Many non-invasive transcutaneous pacemakers operate
Pacing Readiness 6
in two modes: fixed rate (also referred to as asynchronous
Immediate Emergent Pacing
or non-demand) and demand mode (also referred to as synchronous).
Fixed Rate Mode Pacing
Hemodynamically symptomatic, compromising bradycardias that are too slow and unresponsive to
In the fixed rate mode, the pace rate is set by the clinician
atropine.a Symptoms can include blood pressure less
regardless of the patient's intrinsic heart rate. This option
than 80 mmHg systolic, change in mental status,
is preferable when the ECG signal becomes extremely
angina, pulmonary edema.
noisy due to motion artifact or when the pacemaker is
sometimes unable to sense the intrinsic beat.
Bradycardia with escape rhythms unresponsive to
Another reason for using the fixed rate mode is to
pharmacologic therapy
terminate tachyarrhythmias by overdriving the patient's
Pacing for patients in cardiac arrest with profound
intrinsic beat. This method has been successful in a
bradycardia or PEAb due to drug overdose, acidosis,
limited number of patients7,8,9. The above methods are
or electrolyte abnormalities
not widely used in a clinical setting. The danger with
Standby pacing: prepare for pacing for specific
fixed rate mode pacing is the possibility of further
AMI-associatedb rhythms:
exacerbating the tachyarrhythmia and triggering ventricular fibrillation9. Fixed rate mode does not sense
• Symptomatic sinus node dysfunction• Mobitz type II second-degree heart block
the QRS complex for the R-wave and may pace on the
• Third degree heart block c
T-wave, which could trigger ventricular fibrillation.
• Newly acquired left, right, or alternating BBBb or
bifascicular block
Demand Mode Pacing
In demand mode pacing, the pacer senses the patient's intrinsic heart rate and will pace if the intrinsic signal is
Override pacing of either supraventricular or ventricular tachycardia that is refractory to
slower than the rate programmed by the clinician.
pharmacologic therapy or electrical cardioversion
For example, if the patient's heart rate becomes slower
Bradyasystolic cardiac arrest
than the prescribed setting, the pacer will send an electrical stimulus. If the pacer senses that the heart rate
a. Including complete heart block, symptomatic second-degree
is faster than the pacing rate, it inhibits an electrical
heart block, symptomatic sick sinus syndrome, drug-induced bradycardias (i.e., amiodarone, digoxin, -blockers, calcium
signal. The advantages of demand mode pacing are:
channel blockers, procainamide), permanent pacemaker
competition between the pacemaker stimuli and the
failure, idioventricular bradycardias, symptomatic atrial
intrinsic heart rate is minimized, decreasing the risk of R
fibrillation with slow ventricular response, refractory
on T, and the number of pace pulses introduced are
bradycardia during resuscitation of hypovolemic shock, and bradyarrhythmias with malignant ventricular escape
minimized reducing patient discomfort. For this reason,
demand mode pacing is primarily used as the default setting.
b. PEA indicates pulseless electrical activity;
AMI, acute myocardail infarction; and
During demand mode pacing, Philips' defibrillators
BBB, bundle branch block.
detect R waves, or beats. Intrinsic beats are defined as
c. Relatively asymptomatic second- or third-degree heart block
those that are generated naturally by the patient. Paced
can occur in patients with an inferior myocardial infarction. In
(or captured) beats are defined as those that are a result of
such patients pacing should be based on symptoms of
delivered pacing energy.
Philips' ALS defibrillators also define the paced refractory period, which is simply a period of time after the delivery of a pace pulse. The refractory period is approximately
Non-Invasive Transcutaneous Pacing
340 ms for pacing rates less than or equal to 80 pulses per
algorithm to interpret R-waves, such as tall P-waves, tall
minute (PPM), and approximately 240 ms for pacing
T-waves, aberrantly conducted beats, and P-waves
rates greater than or equal to 90 PPM.
without a conducted R-wave that follows. In most
With the pacer on, the defibrillator marks intrinsic beats
instances, tall T-waves and P-waves can be addressed by
on the R-wave with an arrow on the Efficia DFM100.
selecting different leads to minimize the P-wave or
The intrinsic beats are also marked in the print strip.
T-wave amplitudes relative to the R-wave amplitude.
When an intrinsic beat is detected, the time interval for
You may choose to display an annotated ECG with
the next pace pulse starts at the intrinsic beat. If no
ST/AR arrhythmia beat labels in Wave Sector 2. The
intrinsic beat is detected, the time interval for the next
same ECG source appearing in Wave Sector 1 may be
pace pulse starts at the last pace pulse.
displayed with a six-second delay along with white
Philips' defibrillators use the ST/AR Basic Arrhythmia
arrhythmia beat labels. The R-waves marked in Wave
Algorithm for arrhythmia analysis. This analysis provides
Sector 1 are annotating the intrinsic beats that control
information on your patient's condition, including heart
the demand pacing function. The R-waves marked with
rate and arrhythmia alarms. The lead appearing in Wave
beat labels in Wave Sector 2 control the analysis and
Sector 1 is used for arrhythmia analysis. Because the
arrhythmia alarms.
ST/AR Basic Arrhythmia Algorithm is the source needed
Clinicians must not rely solely on the defibrillator's
to generate heart rate and heart rate alarms, the algorithm
classification of beats as intrinsic or paced to determine
cannot be disabled. However, if desired, arrhythmia and
electrical capture. Consider the situation where the
heart rate alarms can be turned off. During arrhythmia
patient's intrinsic HR is 62, and the pacer is set at a rate
analysis, the ECG monitoring parameter continuously:
of 60. Since the two rates are very close, pacer spikes and
Optimize ECG signal quality to facilitate arrhythmia
intrinsic beats may occur very close to each other for
analysis. The ECG signal is continuously filtered to
several seconds. In this circumstance, the defibrillator
remove baseline wander, muscle artifact, and signal
may think the beats are paced based on its simple timing
algorithm, but in fact the beats are intrinsic and the timing coincidental. There may also be cardiac
Measure signal features such as R-wave height, width,
conditions which can cause a truly paced beat to fall
outside of the refractory period.
Create beat templates and classify beats to aid in
Capture must be determined by a clinician who is trained
rhythm analysis and alarm detection.
to interpret the ECG being displayed by the defibrillator.
Examine the ECG signal for ventricular arrhythmias
The defibrillator's marking of beats can be used as a
and asystole.
guide, but not as an absolute indication, of whether or
It is impossible to design a computerized arrhythmia
not capture has been achieved.
algorithm that accurately identifies R-waves 100% of the time. Several conditions can cause difficulty for the
How to Implement Non-Invasive Pacing
Pads Application and Placement
the pads since reversing electrode pad placement increases the pacing threshold. Thus, more current will
Before applying the pads, clean and dry skin sites with a
be needed to capture the heart resulting in greater patient
towel. It may be necessary to shave or clip excessive hair
discomfort10. The most common electrode placement is
in the area of the pads. Follow the manufacturer's
the anterior-anterior position ).
suggested placement of electrode pads. Do not reverse
How to Implement Non-Invasive Pacing
be viewed and exported to a data card, and the date and
Anterior-Anterior Position
time may be changed. However, a password is required to change the configuration of the device. When modifying configuration settings, the device should be connected to external power and have a battery with at least 20% capacity installed.
Changing Rate and Output (mA) During an Event
The pacer rate can be adjusted during pacing from the
default configuration by pressing the [Pacer Rate] soft
The anterior-posterior position is also an effective
key on the device and then the navigational keys to select
the desired number of pace pulses per minute. The pacer output can also be adjusted during pacing from the
Anterior-Posterior Position
default configuration by pressing the [Pacer Output] soft
key and then the navigational keys to:
a. Increase the output until electrical capture occurs. Electrical capture is indicated by the appearance of a QRS complex after each pacing marker.
b. Decrease the output to the lowest level that still maintains capture.
Presence of a peripheral pulse must be verified to assess mechanical capture.
During pacing, verify that arrows appear above the R-wave on the ECG waveform. A single arrow should be
Monitoring Electrodes
associated with each R-wave. If the arrows do not appear or do not coincide with the R-wave, select another lead.
When performing demand mode pacing, the patient
Press the Lead Select button to select the best lead with
must be monitored through either 3- or 5-lead
an easily detectable R-wave. If you are using
monitoring electrodes. Philips' ALS defibrillator/
anterior-anterior pads placement while pacing and are
monitors use the heart rate from this monitoring source
experiencing difficulty with Lead II, select another lead.
to determine if a paced pulse should be delivered.
If a lead cannot be found with correct markers or arrows
Monitoring electrodes are not required for fixed mode
above the R-waves, then switch to fixed mode pacing.
Changing to Fixed Mode Pacing During an
Setting Pacing Rate and Pacing
With the Efficia DFM100, you can switch to fixed mode pacing during an event, as follows:
Setting the Default Configuration
Press the [Pause Pacing] soft key and acknowledge the
The Efficia DFM100 allows you to customize the
defibrillator to best meet your needs with configurable
Press the Smart Select knob and select Pacer Mode.
default pacer rate and output settings. Configuration is
Use the Smart Select knob to highlight Fixed and press
performed through Configuration mode of the device
the Menu Select button.
and may be saved to a data card for replication on multiple devices. At any time, configuration settings may
Press the [Resume Pacing] soft key to resume pacing in
Non-Invasive Transcutaneous Pacing
What to Look for When Pacing
Therefore, a patient's response to pacing must be verified by signs of improved cardiac output, such as: a palpable
pulse rate the same as the rate which pace pulses are being delivered, a rise in blood pressure, and/or improved skin
Capture is defined as gaining control of the patient's
dysfunctional heart by an electrical stimulus (that is, depolarization). Electrical capture can be determined by
An integral or standalone pulse oximeter can be useful for
observing the monitor, which should show a clear
confirming capture (by comparing the pulse rate
indication of the ECG and the pulse marker. Skeletal
measured by the pulse oximeter to the set pacing rate)
muscle contraction is not an indication that capture has
and perfusion (by measuring blood oxygen saturation,
been established, nor is electrical capture alone an
SpO2)11. Philips' defibrillator/monitors integrate the
indication of effective cardiac perfusion. The patient may
optional SpO2 package into the defibrillator for speed,
be suffering from pulseless electrical activity (PEA),
simplicity, and convenience. The SpO2 percentage (%)
previously referred to as electromechanical dissociation
and SpO2 alarm violations will be recorded in the Event
(EMD). When the patient is successfully paced:
electrical and mechanical capture are achieved
Another important factor in effective pacing is the amount of current applied. Sufficient current must be
cardiac output may improve
applied to stimulate the heart into contractions. Unlike
the patient's pulse rate is at least equal to the pacer
invasive pacemakers, non-invasive pacemakers must pass
electrical current through the skin and thorax. This
requires significantly higher current. The higher current
blood pressure may improve
loads can cause the chest muscles to contract and relax
skin color may improve
strenuously, which may be painful and distressing for the
A pulse oximeter is a powerful tool when used in
conjunction with the pacer/ECG monitor for confirming
Using the lowest output setting necessary to achieve
capture because both mechanical and electrical activities
capture will minimize the discomfort. If the discomfort is
can be measured. ECRI provided a theoretical and an
intolerable, sedation and/or analgesia may be necessary.
operational overview of non-invasive pacemakers and concluded that monitoring ECG alone is not enough to
Patient Discomfort
verify that the patient's heart is providing cardiac output.
Pacing will likely cause some discomfort in conscious
The article states that:
patients. Two types of discomfort often experienced are a
Clear recognition of capture continues to be a
burning sensation at the electrode site, and muscle
challenge with today's non-invasive pacemakers.
contraction. Thus, setting the expectations with the patient and family is very important. This should
Artifact from skeletal muscle contractions induced by
alleviate any anxiety or tension and allow the patient to
pacing stimuli can mimic a captured waveform,
become more relaxed. Patients should always be
making verification of capture with the ECG alone
comforted throughout the procedure, and sedation may
be necessary. Make sure the pads are positioned according
Electrical capture does not always produce an effective
to the manufacturer's recommendation. Avoid placing
mechanical (cardiac) contraction.
the pads directly over lesions, injuries or large bone structures (sternum, spine, scapula). Otherwise, more pacer output will be needed to achieve capture.
Special Pacing Situations
Special Pacing Situations
Pediatric Pacing
Atrial flutter is faster in children than in adults.
Newborns may have rates of 600 BPM with distinct
If transcutaneous pacing is elected, the physiological
P-waves. The usual ventricular rate in a newborn with
differences between adults and pediatric patients must be
atrial flutter is 200 to 300 BPM12.
Ventricular tachycardia in infants under the age of
Electrode Size per AAMI Specification
two years averages 260 BPM, which is similar to the rates for supraventricular tachycardia13.
adult — total active area of both pads combined
should be at least 150 cm2. Adult pads are
recommended for use on children weighing 10 kg or
The discomfort associated with pacing may necessitate
sedation in younger conscious children.
pediatric — total active area of both pads combined
should be at least 45 cm2. Recommended for use on
children weighing less than 10kg.
The use of Lead II is not recommended when using
See ANSI/AAMI DF2-1996.
anterior-anterior lead placement.
NOTE: Unlike adult pacing, the use of pediatric
transcutaneous pacing will likely be limited for
Electrodes should be checked frequently (every thirty
cardiopulmonary arrest since most complications are
minutes) to avoid burns and should not be left on for
precipitated by respiratory failure rather than by a
more than two hours.
primary cardiac problem14.
Higher Heart Rate
Defibrillation During Pacing
100 BPM is considered tachycardia in adults. A heart
In defibrillator/pacer combinations, Philips defibrillators
rate less than 100 BPM is considered bradycardia in a
will automatically terminate pacing once the defibrillator
charge switch is enabled.
Supraventricular tachycardia is faster in children than
in adults (the heart rate is typically 300 BPM in infants younger than three months of age). Supraventricular tachycardia is more common than ventricular tachycardia in children.
Non-Invasive Transcutaneous Pacing
Troubleshooting Transcutaneous Pacing
Is there discomfort or pain during pacing?
Check electrode/patient contact.
Explain and set expectations with the patient on
Check electrode connections.
the procedure. Comfort and provide
ECG signal may be too noisy (see below).
encouragement to the patient.
Is there a noisy ECG signal?
Make sure electrodes have been applied following
manufacturer's guidelines. Electrodes should not
Was the patient's skin clean and dry and excessive
be placed over bone (i.e. on the sternum or
hair removed prior to electrode attachment?
Was the expiration date checked before applying
Consider analgesia or sedation.
the electrodes? Dry gel will have an adverse effect on the ECG signal.
Is there electrical capture?
Check the electrode connections.
Pacer output may be too low. Increase pacer
Were the electrodes positioned according to the
Observe the ECG monitor and not just the
patient. Electrical capture occurs when there is a
Is there redness or burns after electrode removal?
consistent ST segment and T-wave after each
Ring-shaped erythema can be expected after
pacing. The appearance is similar to the erythema
Check electrode placement.
seen after receiving shocks from standard external paddle sets. The reaction may vary from patient to
Check contact between the electrode and the
patient due to a number of variables: skin
composition, medication at the time of pacing,
Check to see if the pacer is turned on.
age (very young and older patients may be more
Check viability of the patient.
prone to erythema), dehydration, length of time the patient was paced, and improper pads
Is there mechanical capture?
application. Case reports of burns on neonates and
Check pulse (palpate).
children have been documented. In most cases,
the children had numerous other complications
Check pulse oximeter (SpO2).
which may have contributed to the burns.
Check blood pressure.
Neonates in particular may be more susceptible to
Check skin color.
thermal injury than adults due to thinner skin, less hair, weaker intercellular attachments, fewer
Evaluate mental status of patient.
eccrine and sebaceous gland secretions, and an
Is the pacer not sensing heart rate in demand mode?
increased susceptibility to external irritants. To
minimize thermal injury, pacing duration should
Check size of R-wave. If it is too small, increase
be kept to a minimum with frequent skin
the ECG size.
inspection (every 30 minutes)15.
Check to see if the pacer is turned on.
Rudd M. Basic concepts of cardiovascular physiology. Hewlett
ECRI. Reprinted with permission from Health Devices,
Packard Company, 1973, Mar; 7:1-8:4.
American Heart Association, ACLS Provider Manual, 2002
Garson A. "Pediatric arrhythmias. How different from
Zoll P.M., "Non-invasive temporary cardiac pacing."
adults?" Med. Surg. Ped., 1987; 9: 543-52.
Journal of Electrophysiology, 1987; 1:156-161.
Garson A., Smith R.T., Cooley D.A., et al. "Incessant
Luck J.C., Davis D. "Termination of sustained tachycardia
ventricular tachycardia in infants: Purkinje hamartomas
by external non-invasive pacing." PACE, 1987;
and surgical cure". J Am Col Card, publication pending.
10 Beland M.J., Hesslein P.S., Finlay C.D., et al.
Barold S.S., Falkoff M.D., Ong L.S., et al. "Termination of
"Non-invasive transcutaneous cardiac pacing in children".
ventricular tachycardia by transcutaneous cardiac pacing."
PACE, 1987 Nov-Dec; 10:1262-70.
American Heart Journal, 1987; 114:180-182.
11 Pride H.B., McKinley D.F. "Third degree burns from the
Bartecchi C.E. "Temporary cardiac pacing — Diagnostic
use of an external cardiac pacing device." Critical Care
and therapeutic indications." Postgraduate Medicine, 1987
Medicine, 1990; 18(5):572-73.
June; 81(8):269-77.
Non-Invasive Transcutaneous Pacing
2014 Koninklijke Philips N.V.
Philips Healthcare is part
All rights are reserved.
Reproduction in whole or in part is prohibited without the prior
On the web
written consent of the copyright
By e-mail
Philips Healthcare reserves the right
to make changes in specifications or
to discontinue any product at any
time without notice or obligation
By postal service
and will not be liable for any
Philips Healthcare
consequences resulting from the use
3000 Minuteman Road
of this publication.
Andover, MA 01810-1085
Asia
Tel: +852 2821 5888
Published March 2014, Edition 1
Europe, Middle East, and Africa
Tel: +49 7031 463 2254
Latin America
Tel: +55 11 2125 0744
Source: http://www.yms.co.za/wp-content/uploads/2016/07/Pacing_App_Note.pdf-nodeid10747034vernum-2.pdf
Principles and processes in biotechnology.pmd
Ever since the days of Rene Descartes, the French philosopher,mathematician and biologist of seventeenth century, all human Chapter 11 knowledge especially natural sciences were directed to develop Biotechnology : Principles and technologies which add to the creature comforts of human lives, as also value to human life. The whole approach to Chapter 12
whatsupdoc.tv
El ejército que ¿Qué queda de las Fuerzas Armadas que rompieron la democracia en Chilehace 30 años? Seguramente, más la forma que el fondo. Por primera vez, la Escue-la Militar, el gran centro de formación de oficiales, ha abierto sus puertas. Asíson los nietos uniformados del ex dictador. Por Francesc Relea. Fotografía de Luis Poirot.